ORP/pH Pre-amplifier

 

The economical Sensorex PHAMP-1 is a battery powered in-line design. It is designed to boost the signal and lower the impedance from a ORP electrode to a ORP monitor, controller or transmitter.  It can be used for long cable runs of up to 1,200 feet (300 meters). Use this preamplifier when your instrument is not designed to power a remote amplifier. The PHAMP-1 has an operational output of +/- 2000 mV.

click image to enlarge

pH/ORP Electrode Amplifier (PHAMP-1)
Product Features
PHAMP1  +/- 2000 mV
  Est. 3 year Battery Life
REDOX/ORP Electrode Pre-Amplifier Specifications

 

The PHAMP-1 can also be used to input a ORP electrode signal directly into a device which can accept voltage inputs. The batteries are sealed, and have an estimated life of more than 3 years. The BNC connector from your coax electrode cable is attached to the BNC Receptacle INPUT Connection on the PHAMP-1. The output is a BNC connector. You must have a BNR receptacle on your Cable to connect from the output and go to your controller.  Sensorex provides accessory cables and adapters to make this transition. The PHAMP-1 is not water proof, you must protect it from water and the elements.

For ORP Calibration/Verification solutions, see our ORP buffers page for complete details.