|
DO Probes
Dissolved
Oxygen (DO) is the term used for the measurement of the amount
of oxygen dissolved in a unit volume of water. In water quality
applications, such as aquaculture (including fish farming) and
waste water treatment, the level of DO must be kept high. For
aquaculture if the DO level falls too low the fish will suffocate.
In sewage treatment, bacteria decompose the solids. If the DO
level is too low, the bacteria will die and decomposition ceases;
if the DO level is too high, energy is wasted in the aeration
of the water. With industrial applications including boilers,
the make-up water must have low DO levels to prevent corrosion
and boiler scale build-up which inhibits heat transfer.
Although dissolved oxygen
(DO) is usually displayed as mg/L or ppm, DO sensors do
not measure the actual amount of oxygen in water, but instead
measure partial pressure of oxygen in water. Oxygen pressure is
dependent on both salinity and temperature.
There are
two fundamental techniques for measuring DO— galvanic and
polarographic. Both probes use an electrode system where the DO
reacts with the cathode to produce a current. If the electrode
materials are selected so that the difference in potential is
-.5 volts or greater between the cathode and anode, an external
potential is not required and the system is called galvanic. If
an external voltage is applied, the system is called polarographic.
- Galvanic probes are more stable and more accurate at low dissolved
oxygen levels than polarographic probes.
- Galvanic probes often operate several months without electrolyte
or membrane replacement, resulting in lower maintenance cost.
- Polarographic probes need to be recharged every several weeks
of heavy use.
Galvanic DO sensors consist
of two electrodes, an anode and cathode which are both immersed
in electrolyte (inside the sensor body). An oxygen permeable membrane
separates the anode and cathode from the water being measured.
Oxygen diffuses across the membrane. It interacts with
the probe internals to produce an electrical current (more detail
is shown below the DO sensor graphic). Higher
pressure allows more oxygen to diffuse across the membrane and
more current to be produced. The actual output from the
sensor is in millivolts. This is achieved by passing the
current across a thermistor (a resistor that changes output with
temperature).
V = i *
R,
V is output
in Volts, i = current
R is resistance
from thermistor in ohms
The thermistor corrects
for membrane permeability errors due to temperature change.
In other words, increasing permeability at higher temperature
allows more oxygen to diffuse into the sensor, even though the
oxygen pressure has not changed. This would give falsely
high DO if the thermistor were not used.
To represent sensor output in ppm
or mg/L, the temperature of the water must be known. A
separate temperature sensor can be used or one can built into
the sensor. This is independent from the thermistor connected
between the anode and cathode to compensate for membrane permeability
changes due to temperature change.

Some characteristics
of membrane DO probes are:
- The pH of the solution does not affect the performance of
membrane probes.
- Chlorine and hydrogen sulfide(H2S) cause erroneous readings
in DO probes.
- Atmospheric pressure (altitude above sea level) affects the
saturation of oxygen. DO probes must be calibrated for the barometric
pressure when reading in mg/l (ppm).
- Membrane thickness determines the output level of the probe
and the speed of response to change in DO levels.
- Salinity correction must be made
The cathode
is a hydrogen electrode and carries a negative potential with
respect to the anode. Electrolyte surrounds the electrode pair
and is contained by the membrane. With no oxygen, the cathode
becomes polarized with hydrogen and resists the flow of current.
When oxygen passes through the membrane, the cathode is depolarized
and electrons are consumed. The cathode electrochemically reduces
the oxygen to hydroxyl ions:
O2
+ 2 H2O + 4 e- = 4 OH-
The anode
reacts with the product of the depolarization with a corresponding
release of electrons.
Zn
+ 4 OH- = Zn(OH)42- + 2e-
The electrode
pair permits current to flow in direct proportion to the amount
of oxygen entering the system. The magnitude of the current gives
us a direct measure of the amount of oxygen entering the probe.
Because
all of the oxygen entering the probe is chemically consumed, the
partial pressure of oxygen in the electrolyte is zero. Therefore,
a partial pressure gradient exists across the membrane and the
rate of oxygen entering the probe is a function of the partial
pressure of oxygen in the air or water being measured.
Since the
partial pressure of dissolved oxygen is a function of temperature
of the sample, the probe must be calibrated at the sample temperature
or the probe’s meter must automatically compensate for varying
sample temperature. Note that this thermal effect is different
from the thermal response of the membrane discussed above.
The reading
of a DO probe must be corrected for the amount of salt in the
sample. As seen in the chart below, the salt in solution will
reduce the actual concentration of oxygen.
In
all DO Probes, the membrane/sample interface should have a few
cm/sec flow of the sample for precision performance. Without flow
at the interface, the surrounding oxygen will be consumed and
the local reading drops. The output of the probe increases(up
to a point) with relative movement between the probe and sample.
Dissolved
Oxygen Measurement
The amount
of oxygen that a given volume of water can hold is a function
of the atmospheric pressure at the water-air interface; the temperature
of the water; and the amount of other dissolved substances (such
as salts or other gases) in the water. Recall seeing bubbles in
a pot of water just before it starts to boil. These bubbles are
the air which was dissolved in the water at room temperature.
When the water boils, the dissolved oxygen is ejected—warmer
water contains less DO. When other substances, such as salts,
are dissolved in a unit volume of water, there is less room for
oxygen to dissolve—oxygen is less soluble than most salts
The following
table shows the relationship of dissolved oxygen (mg/L) to temperature
and salinity:
Oxygen Saturation Based on Temperature and Salinity
| 0 |
14.62ppm |
13.73 |
12.89 |
12.10 |
11.36 |
10.66 |
| 10 |
11.29 |
10.66 |
10.06 |
9.49 |
8.96 |
8.45 |
| 20 |
9.09 |
8.62 |
8.17 |
7.75 |
7.35 |
6.96 |
| 25 |
8.26 |
7.85 |
7.46 |
7.08 |
6.72 |
6.39 |
| 30 |
7.56 |
7.19 |
6.85 |
6.51 |
6.20 |
5.90 |
| 40 |
6.41 |
6.12 |
5.84 |
5.58 |
5.32 |
5.08 |
The
relationship between temperature, salinity, and dissolved oxygen
is approximated with the following exponential equation:
ln(
C ) = -139.34 + (1.5757 x 105/T) - (6.6432 x 107/T2) + (1.2438
x 1010/T3)
-
(8.6219 x 1011/T4) - S [1.7674 x 10-2 - (10.754/T) + (2.1407 x
103/T2)]
T = Temperature in degree
Kelvin
S = Salinity in parts
per thousand (ppt)
C = Concentration in
mg/L
As the
pressure of the air above the water is increased, more oxygen
will be dissolved in the water. This increases the concentration
of the dissolved oxygen. The solubility of a gas in a liquid is
directly proportional to the pressure of that gas above the liquid—Henry’s
law. This is often expressed as:
p
= k C (C = concentration of DO)
If different
gases are mixed in a confined space of constant volume and at
a definite temperature, each gas exerts the same pressure as if
it alone occupied the space. The pressure of the mixture as a
whole is the total of the individual or partial pressure of the
gases composing the mixture—Dalton’s law of partial
pressures. The partial pressure of each gas is proportional to
the number of molecules of that gas in the mixture. Air is 20.948%
oxygen. When air bubbled through water, only 20% as much oxygen
dissolves as would dissolve if pure oxygen were used instead of
air, at the same pressure.
Concentration
of dissolved oxygen is also measured in units of % saturation.
% saturation is simply the ratio of the measured mg/L of dissolved
oxygen divided by the mg/L of dissolved oxygen at saturation—as
given in the above tables, saturation levels is dependent upon
the temperature, salinity, and pressure. Since % saturation is
a ratio, it is not affected by these conditions if the calibration
at 100% saturation was performed under the same conditions.
Solubility
of solutes as a function of temperature
(mg
of solutes per liter of water):
| 02 |
69 |
43 |
31 |
14 |
0 |
|
| CO2 |
3350 |
1690 |
970 |
580 |
|
|
| NaCl |
357,000 |
360,000 |
366,000 |
373,000 |
384,000 |
398,000 |
| KCl |
276,000 |
340,000 |
400,000 |
455,000 |
511,000 |
567,00 |
With
stationary, continuously monitoring Dissolved Oxygen Probes, the
source of the oxygen being measured is air. Thus, Dissolved Oxygen
in air or saturated water (mg/l or ppm) as a function of temperature
is determined by:
Solubility
(ml/L) x Density (mg/ml) x % in air = saturated DO in mg/L (ppm)
Solubility
(mg/L) x % in air = saturated DO in mg/L (ppm) Increasing temperature usually
increases the solubility of solids and liquids whereas it reduces
the solubility of gases. Also keep your units straight--mg/L,
ppm, ml/L, % saturation. |