|
What Is pH?
pH is a
measure of acidity or alkalinity. The amount of hydrogen ions
(H+) causes a liquid to be acidic (high concentration of hydrogen
ions) or alkaline (low concentration of hydrogen ions).
The pH range is measured from 0 to 14. Values below 0 or above
14 are possible but rare and cannot be measured with our electrodes.
The pH scale is derived from the dissociation constant of water
in the following equation:
H2O
-> H+ + OH- = 1 x 10-14 (mol/L)2 = Kw
(Kw
is the dissociation constant of water).

Hydrogen Ion Concentration in Moles/Liter at 25° C
| 0 |
1.0 |
0.00000000000001 |
| 1 |
0.1 |
0.0000000000001 |
| 2 |
0.01 |
0.000000000001 |
| 3 |
0.001 |
0.00000000001 |
| 4 |
0.0001 |
0.0000000001 |
| 5 |
0.00001 |
0.000000001 |
| 6 |
0.000001 |
0.00000001 |
| 7 |
0.0000001 |
0.0000001 |
| 8 |
0.00000001 |
0.0000001 |
| 9 |
0.000000001 |
0.00001 |
| 10 |
0.0000000001 |
0.0001 |
| 11 |
0.00000000001 |
0.001 |
| 12 |
0.000000000001 |
0.01 |
| 13 |
0.0000000000001 |
0.1 |
| 14 |
0.00000000000001 |
1.0 |
pH of Common Materials

Why
is pH an Important Measurement?
It determines
product quality in:
Sugar Refining
Pulp and Paper Mills
Latex Coagulation
Photo Developing
It enhances product efficiency of:
Flue gas scrubbing
Circuit Board Etching
Fermentation processes
It assures product safety
Chromate and cyanide destruct systems
Potable and waste waters
Food, low pH to prevent botulism
How
Can pH be Measured?
Colorimetric
Methods
- Reagent addition
- pH paper
Electrochemical
Methods (pH electrodes)
A pH Measuring System
Consists of:
- A pH electrode, an electrode whose
output voltage changes as the pH (hydrogen ion concentration)
changes
- A reference electrode is an electrode
whose voltage output stays constant
- A pH meter, a millivolt meter with
a special high impedance input circuit and circuits to change
the electrode's millivolts into pH unit readouts.
- Optionally, an automatic temperature compensator,
a device which senses temperature so that the meter can correct
for the effects of temperature changes.
How
Does A pH Electrode Work?
- Special composition glass senses
H+ and a millivoltage is
generated (59.2 mV per pH unit at 25C.)
- A filling solution picks up the signal from the special pH
glass
- A pure silver wire dipped in silver chloride passes the signal
from the solution whose pH is being measured to the electrode's
cable or connector
- Sodium Ion error for solutions > pH 12.3
How
Does A Reference Electrode Work?
- A porous reference junction separates the filling solution
in the electrode from the solution whose pH is to be measured.
- The filling solution's constant chloride ion concentration
generates a millivoltage at a pure silver wire with silver chloride
on it.
- The silver wire passes the signal from the solution being
measured to the electrode's cable or connector.

Single
vs. Double Junction Reference
Chemicals that cause silver to precipitate at the reference junction
will contaminate and plug single junctions. These may be such
compounds as sulfides, mercaptans, cyanides, Iodides, and proteins.
Other compounds such as silver, lead, mercury, and other heavy
metal compounds will react with the chloride in the gel, causing
a reduction in the reference output. Selection of the proper chemistry
in the lower (double) junction will prevent or at least minimize
the negative effects of these reactive compounds.
How
Does A Combination Electrode Work?
A combination electrode
consists of a pH electrode and a reference electrode built into
a single body or housing. A combination electrode therefore
works like the pH and reference electrodes combined!
How
does a pH meter Work?
A pH meter
takes the input from the pH glass (high impedance mV) and the
input from the reference sensor and compares these 2 values to
get a resulting millivolt reading. The reading in mV is
converted to pH by the following guidleine:
Zero mV = pH 7
59.2mV per pH unit change
mV are + for pH <7 and mV are - for pH >7
Example of pH Circuit

Note that the pH and
reference input go into an operational amplifier (op amp) due
to the very high resistance of the pH glass. The meter will
also adjust zero and span offsets and can do automatic temperature
compensation for pH error (discussed below).
Temperature
Compensation
When measuring pH using
a pH electrode the temperature error from the electrode varies
based on the Nernst Equation as 0.03pH/10C/unit of pH away from
pH7. As shown in the table below, the error due to
temperature is a function of both temperature and the pH
being measured. Note that there is
no error at pH7 and 25 Deg C. Temperature compensation
can be achieved manually or automatically. Manual temperature
compensation is usually achieved by entering the temperature of
the fluid being measured into the instruments menu and then the
instrument will display a "Temperature Compensated"
pH reading. This means that the temperature is corrected
to the value expected at 25 Deg C. Automatic temperature
compensation requires input from a temperature sensor and constantly
sends a compensated pH signal to the display. Automatic
temperature compensation is useful for measuring pH in systems
with wide variations in temperature.
pH vs. Temperature Error Chart
| |
pH 2 |
pH 3 |
pH 4 |
pH 5 |
pH 6 |
pH 7 |
pH 8 |
pH 9 |
pH 10 |
pH 11 |
pH 12 |
| 5° |
.30 |
.24 |
.18 |
.12 |
.06 |
0 |
.06 |
.12 |
.18 |
.24 |
.30 |
| 15° |
.15 |
.12 |
.09 |
.06 |
.03 |
0 |
.03 |
.06 |
.09 |
.12 |
.15 |
| 25° |
0 |
0 |
0 |
0 |
0 |
0 |
0 |
0 |
0 |
0 |
0 |
| 35° |
.15 |
.12 |
.09 |
.06 |
.03 |
0 |
.03 |
.06 |
.09 |
.12 |
.15 |
| 45° |
.30 |
.24 |
.18 |
.12 |
.06 |
0 |
.06 |
.12 |
.18 |
.24 |
.30 |
| 55° |
.45 |
.36 |
.27 |
.18 |
.09 |
0 |
.09 |
.18 |
.27 |
.36 |
.45 |
| 65° |
.60 |
.48 |
.36 |
.24 |
.12 |
0 |
.12 |
.24 |
.36 |
.48 |
.60 |
| 75° |
.75 |
.60 |
.45 |
.30 |
.15 |
0 |
.15 |
.30 |
.45 |
.60 |
.75 |
| 85° |
.90 |
.72 |
.54 |
.36 |
.18 |
0 |
.18 |
.36 |
.54 |
.72 |
.90 |
Note:
Values in light blue are less than .1 error and may not require
temperature compensation. Values in gray are temperature
and pH in which there is no error in pH from temperature.
How
are pH Systems Calibrated?
- The electrodes are placed in buffers. Buffers are solutions
of know, stable pH value
The commonly used buffers have pH values of 4.01, 7.00 and 10.00
The pH values of buffers change with changes of temperature
(and so does the pH of solutions being measured)
- First the the system's (electrode and meter together) Zero
Point is adjusted
Zero point is usually determined with
pH 7.00 buffer
Ideally, a buffer is used whose value is close to that of the
material or solution being measured
A "calibrate "or "standardize" adjustment
potentiometer knob or push button on the pH meter is used to
set the system (electrode and meter together) to read the buffer's
pH value
- Next, the system's span is checked and/or adjusted. The electrodes
are rinsed and put into a second buffer
The system should read close to the pH value of the second buffer.
Most meters have controls labeled "SPAN" or "SLOPE"
which compensates for electrodes with spans that are too short.
- The frequency of calibration is at the discretion of the user.
- NOTE: 2-point calibration is necessary to ensure
electrode works properly since a broken electrode can give acceptable
pH7 output in calibration mode.
pH
Applications
Wastewater neutralization
Electroplating
Chemical Manufacturing
Circuit board etching
Flue gas scrubbers
Boilers and cooling towers
Pulp and paper mfg.
Food and Beverage
Pharmaceuticals
inks, paints latex
Water treatment (sewage)
Aquariums, aquaculture
Fermentation (wine, beer, alcohol)
|