Having a problem with your conductivity system or electrode? Here is a list of some common symptoms, causes, and fixes to commonly encountered conductivity problems. We hope this helps!
This presentation is a systematic procedure to isolate problems in your conductivity measuring system.
The following tools are recommended to assist you in problem identification:
The meter or controller is the easiest component to eliminate as a possible cause of your problem. Disconnect conductivity sensor from meter/controller and connect conductivity simulator in its place. Input conductivity readings and make sure controller/meter display matches the reading from the simulator. Use fixed resistor if a simulator is not available. If readings do not match, the problem is in the meter or controller. Contact meter/controller supplier for help or refer to meter/controller troubleshooting section. If simulator inputs are correctly displayed on the meter/controller, then the electrode is the most likely cause of the problem.
Check electrode in conductivity calibration/standard solutions as recommended above. Compare results to those on the following table:
| Conductivity standard solution output | Possible Cause | Corrective Action |
| No output |
a) Conductivity sensor not connected to instrument b) Bad connection at connector c) Break in cable or internal connection broken in sensor |
a) Check all connections from electrode to instrument b) Check for electrical continuity from conductivity pin to connector using Ohms setting on DVM (should be <1 Ohm) c) Contact Sensorex |
| Output in standard solution is more than 10% different than value of solution |
a) Solution is not fresh, contaminated, or labeled wrong b) Electrode is dirty c) Standard solution is outside of cell range** |
a) Use fresh solutions or new lost of solution b) Clean electrode per instruction manual. Do not abrade measuring surface. c) Pick standards based on ranges shown below** |
| Unstable or drifting reading | Bubbles in electrode or flow cell |
a) Shake or stir electrode to remove bubbles if testing solution in beaker. b) Mount electrode horizontally so bubbles rise up and away from electrode. |
| Readings in line do not change but electrode calibrates in conductivity standard solutions |
a) Electrode not completely immersed in process fluid b) Electrode not connected to controller |
a) Check installation fittings and make sure sensor is completely installed. b) Check wiring to instrument |
| Cell Constant (k) | Conductivity Range(uS) |
| 0.10 | 5-5,000 uS |
| 1.0 | 50-50,000 uS |
| 10.0 | 500-20,000 uS |
If you are still having problems after trying the above diagnostics, fill out our Application Questionnaire and e-mail it to us for assistance.