Having a problem with your DO measurement? See helpful troubleshooting tips below.
This presentation is a systematic procedure to isolate problems in your dissolved oxygen measuring system.
The following tools are recommended to assist you in problem identification:
Check electrode in conductivity calibration/standard solutions as recommended above. Compare results to those on the following table:
| DO Sensor output/Symptom | Possible Cause | Corrective Action |
| No output |
a) DO sensor not connected to instrument b) Bad connection at connector c) Break in cable or internal connection broken in sensor |
a) Check all connections from electrode to instrument b) Check for electrical continuity from conductivity pin to connector using Ohms setting on DVM (should be <1 Ohm) c) Contact Sensorex |
| Submersible probe output does not match handheld DO meter output | a) Probes not calibrated | a) Calibrate both probes before comparing outputs |
| DO reading is unstable and drifts downward | Electrolyte depleted | Replacement membrane and electrolyte per product instructions |