pH is a measure of acidity or alkalinity. The amount of hydrogen ions (H+) causes a liquid to be acidic (high concentration of hydrogen ions) or alkaline (low concentration of hydrogen ions). The pH range is measured from 0 to 14. Values below 0 or above 14 are possible but rare and cannot be measured with our electrodes. The pH scale is derived from the dissociation constant of water in the following equation:
| pH | H+ conc. | OH - conc. |
| 0 | 1.0 | 0.00000000000001 |
| 1 | 0.1 | 0.0000000000001 |
| 2 | 0.01 | 0.000000000001 |
| 3 | 0.001 | 0.00000000001 |
| 4 | 0.0001 | 0.0000000001 |
| 5 | 0.00001 | 0.000000001 |
| 6 | 0.000001 | 0.00000001 |
| 7 | 0.0000001 | 0.0000001 |
| 8 | 0.00000001 | 0.0000001 |
| 9 | 0.000000001 | 0.00001 |
| 10 | 0.0000000001 | 0.0001 |
| 11 | 0.00000000001 | 0.001 |
| 12 | 0.000000000001 | 0.01 |
| 13 | 0.0000000000001 | 0.1 |
| 14 | 0.00000000000001 | 1.0 |
It determines product quality in:
It enhances product efficiency of:
Colorimetric Methods
Electrochemical Methods (pH electrodes)
A pH Measuring System Consists of:
Chemicals that cause silver to precipitate at the reference junction will contaminate and plug single junctions. These may be such compounds as sulfides, mercaptans, cyanides, Iodides, and proteins. Other compounds such as silver, lead, mercury, and other heavy metal compounds will react with the chloride in the gel, causing a reduction in the reference output. Selection of the proper chemistry in the lower (double) junction will prevent or at least minimize the negative effects of these reactive compounds.
A combination pH electrode consists of a pH electrode and a reference electrode built into a single body or housing. A combination electrode therefore works like the pH and reference electrodes combined!
A pH meter takes the input from the pH glass (high impedance mV) and the input from the reference sensor and compares these two values to get a resulting millivolt reading. The reading in mV is converted to pH by the following guidleine:
Note that the pH and reference input go into an operational amplifier (op amp) due to the very high resistance of the pH glass. The meter will also adjust zero and span offsets and can do automatic temperature compensation for pH error (discussed below).
When measuring pH using a pH electrode the temperature error from the electrode varies based on the Nernst Equation as 0.03pH/10C/unit of pH away from pH7. As shown in the table below, the error due to temperature is a function of both temperature and the pH being measured. Note that there is no error at pH7 and 25° C. Temperature compensation can be achieved manually or automatically. Manual temperature compensation is usually achieved by entering the temperature of the fluid being measured into the instruments menu and then the instrument will display a "Temperature Compensated" pH reading. This means that the pH value is corrected to the value expected at 25° C. Automatic temperature compensation requires input from a temperature sensor and constantly sends a compensated pH signal to the display. Automatic temperature compensation is useful for measuring pH in systems with wide variations in temperature.
| pH 2 | pH 3 | pH 4 | pH 5 | pH 6 | pH 7 | pH 8 | pH 9 | pH 10 | pH 11 | pH 12 | |
| 5° | .30 | .24 | .18 | .12 | .06 | 0 | .06 | .12 | .18 | .24 | .30 |
| 15° | .15 | .12 | .09 | .06 | .03 | 0 | .03 | .06 | .09 | .12 | .15 |
| 25° | 0 | 0 | 0 | 0 | 0 | 0 | 0 | 0 | 0 | 0 | 0 |
| 35° | .15 | .12 | .09 | .06 | .03 | 0 | .03 | .06 | .09 | .12 | .15 |
| 45° | .30 | .24 | .18 | .12 | .06 | 0 | .06 | .12 | .18 | .24 | .30 |
| 55° | .45 | .36 | .27 | .18 | .09 | 0 | .09 | .18 | .27 | .36 | .45 |
| 65° | .60 | .48 | .36 | .24 | .12 | 0 | .12 | .24 | .36 | .48 | .60 |
| 75° | .75 | .60 | .45 | .30 | .15 | 0 | .15 | .30 | .45 | .60 | .75 |
| 85° | .90 | .72 | .54 | .36 | .18 | 0 | .18 | .36 | .54 | .72 | .90 |
Note: Values in light blue are less than 0.1 error and may not require temperature compensation. Values in gray are temperature and pH in which there is no error in pH from temperature.