|
This presentation is a systematic
procedure to isolate problems in your pH measuring system.
Troubleshooting
Equipment
The following
tools are recommended to assist you in problem identification:
- Portable pH/mV meter with interconnect
cable
The meter and buffers can verify operational status of a pH electrode
or ORP electrode.
- A DVM (digital volt meter)
If your pH system has Automatic Temperature Compensation (ATC),
the DVM will verify operation of the ATC element. Refer to the
ATC page or your meter's instruction manual for the appropriate
ATC resistance reading (typically 100, 1000, 3K, 10K, or 30K ohms).
- pH Simulator
Simulators similar to Sensorex
pH Checker can quickly identify or eliminate the or controller
as the cause of the problem. Check under high impedance condition--an
option on the pH Checker
- pH
Buffers
Use 7 and either 4 or 10 buffer depending upon at which end of
the pH scale you normally operate in.
Meter,
Preamplifier, Controller Troubleshooting
The meter or controller
is the easiest component to eliminate as a possible cause of your
problem. Follow the stepwise instructions below for quick
problem solving.
- Disconnect the pH electrode from the meter, preamplifier (if
you have an external preamp), or controller.
- Connect a pH Simulator (like the Sensorex pH Checker) in the
position where the electrode was connected.
- Set the pH simulator to 7 with high impedance and verify the
meter, preamplifier, or controller output .
- Reset the pH simulator to 4 with high impedance and verify the
output. If you did not get 7.00 and 4.00 pH, recalibrate or troubleshoot
the meter, preamplifier, or controller in accordance with its
instruction manual. Note that miswired
preamps usually give negative readings or constant zero reading
regardless of electrode or simulator input
pH Electrode Troubleshooting
Remove the
electrode from the sample or process and disconnect it from the
meter or controller. Connect to the portable pH/mV meter and test
electrode in 7 and 4 pH buffers. Compare results to the following
table:
| 6.2-6.8 in all buffers |
a) Cracked pH glass
b) Stress crack |
a) replace electrode
b) Contact Sensorex for Return Authorization |
| 7.00 in all buffers |
a) Bad connection
b) Internal short circuit |
a) check/fix connection
b) Contact Sensorex for Return Authorization |
| Buffers read close to expected value but speed of response*
is slow (>30 seconds) |
a) dirty electrode pH glass and/or reference junction
b) Temperature too low |
a) Clean electrode per instructions included with it when
shipped
b) Flat pH glass pH electrodes should be used at temperature
>10C/50F
Bulb pH should be used at temperature >0C/32F |
| Large offset in buffers** |
a) reference poisoned
b) Ground loop**** |
a) Contact Sensorex about special references
b) Ground solution for tank or line to known earth ground or
buy Sensorex Ground
Loop Interrupt pH electrode |
Short span***
(Typically less than 70% |
a) Dirty pH glass or reference junction
b) aged electrode |
a) clean electrode per electrode's care and use instructions
b) replace electrode (too old) |
| Unstable or drifting reading |
reference dirty or plugged |
clean electrode per electrode's care and use instructions
(do not use abrasives to clean reference
junctions!) |
If you are still having
problems after trying the above diagnostics, fill out our Application
Questionnaire and e-mail it to us for assistance.
Speed
of Response*
New pH electrodes from Sensorex with spherical pH measuring surfaces
will respond to within 95% of their final value in less than 1 second.
New flat surface electrodes (due to their higher initial impedance)
have 95% response in less than 5 seconds. As electrodes age, their
impedance will increase and their Speed of Response will become
slower. Coatings on the pH glass will also give a slower response
time.
Offset*
A perfect electrode will read 7.00 in 7.00 buffer. Offset is defined
as the difference between the electrode's reading in 7.00 buffer
and 7.00. For example, when an electrode reads 7.15 in 7.00 buffer,
its Offset is 0.15 (7.15-7.00 = 0.15). The factory specification
for a new electrode is 7.00 +/-0.20 pH since all modern meters and
controllers are easily capable of correcting such offset.
As electrodes age, their
Offset will change. Usually offset changes are corrected by calibrating
the pH meter or controller with the electrode together. An electrode
is usable as long as the meter or controller can be adjusted to
correct the offset and span and the Speed of Response is acceptable.
Changes in an electrode Offset of more than 0.50 pH in a month or
less may indicate an application problem.
Span***
A perfect electrode will have 100% of the theoretical Span which
means that when an electrode is calibrated to read 7.00 in 7.00
buffer and is place in 4.00 buffer, it will read 4.00. Sensorex
specification for a new electrode is at least 97% of theoretical
Span. So when an electrode is placed in 4.00 buffer, it can read
between 3.90 and 4.10. Similar to Offset, most meters and controllers
have Span adjustments to correct for span errors. Normal electrode
aging will reduce the Span. As long as the pH meter or controller
has sufficient adjustment to correct of the reduced span, the electrode
is usable, providing the speed of response is acceptable to your
process. Coating on the pH sensitive glass can also reduce the Span.
Ground
Loops***
When a pH system is unstable, erratic, or the offset drifts, the
most common problem is an electrical ground loop in the system,
particularly if the tank and/or pipes are plastic. To verify this
problem, remove the electrode and calibrate it in a known buffer
in a beaker. If the electrode measures within specification (stable
and adjustable offset) when calibrated, place a copper wire in the
beaker and the other end in your system. If the reading becomes
unstable or shifts, a ground loop is your problem.
ELECTRODE CALIBRATES
THEN
ELECTRODE READS HIGH
IN BUFFER
OR OUT OF RANGE IN PROCESS

The sources of the ground
loop could be any mixer motor, pump, conductivity probe, or other
electrically powered device in the media with the pH electrode.
Do not attach the conductivity probe or any other electrically powered
device to the same ground on your meter or controller as the pH
electrode. You can also try placing a large (12 or 14 AWG) copper
wire into the media and the other end to the meter or controller
ground terminal to draw the ground loop away from the pH electrode.
Sensorex also offers a Ground
Loop Interrupt electrode to solve this problem. |